Purpose 

The self-test verifies proper machine operation and helps with remote troubleshooting with Klaris support. It runs for about 5-10 minutes. If successful, the power button and progress bar will blink green. 

How to Use 

  • Remove the tray and inserts, then close the lid.
  • Hold down the power button for more than 10 seconds. Once you hear a beep, release the button.
  • The machine will check ambient temperature and ensure all components are at steady-state temperature. This may take 0–5 minutes.
  • The LEDs will cycle through all colors to verify proper light function.
  • The power button and progress bar will then blink white while the machine tests the cooling mechanism, mixing motor, and lid sensor.
  • If an error is detected any time throughout the process, the power button and progress bar will turn red in an on/off/blinking pattern. Match the pattern against the error table in the manual.
  • If the test passes, the power button and progress bar will blink green.
  • To exit, power cycle the machine.